Principle
The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to AFGF/FGF1. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to AFGF/FGF1. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain AFGF/FGF1, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of AFGF/FGF1 in the samples is then determined by comparing the OD of the samples to the standard curve.
Specification
Product Name |
|
|
Cat no | ABQ-ES96-AFGFhu | |
Applications | ELISA | |
Reactivity | Human | |
Alternative Name | FGF-1, ECGF, ECGF-beta, A-FGF, FGFA, ECGFA, ECGFB, FGF-Alpha, HBGF1, HBGF-1, ECGFB, GLIO73, Heparin-binding growth factor 1 | |
Standard | 2000pg/mL | |
Sensitivity | 18.75 pg/mL | |
Detection Range | 31.25-2000pg/mL | |
Sample Type | Serum, Plasma, Tissue homogenate and Other biological samples | |
Sample Volume | 100 μL | |
Assay Type | Sandwich | |
Assay Duration | 3.5H | |
Detection Wavelength | 450nm | |
Storage | 2-8oC/-20 oC |
PRECISION
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Item | Intra-assay Precision | Intra-assay Precision |
Sample | 3 | 3 |
Replicate | 9 | 18 |
CV(%) | 5 | 8 |
Rate of recovery
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type | Range(%) | Average Recovery(%) |
Serum (n=8) | 82-95 | 88 |
EDTA plasma (n=8) | 80-95 | 87 |
Cell culture media (n=8) | 86-99 | 92 |