No products in the cart.
19 people are viewing this product right now
Principle
This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with α-CTx protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to α-CTx. Next,Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of stop solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of α-CTx in the samples is then determined by comparing the OD of the samples to the standard curve.
Specification
Product Name | Quanticon Human α-CTx(Alpha Crosslaps) ELISA Kit |
SKU | ABQ-EC96-ACTXhu |
Applications | ELISA |
Reactivity | Human |
Alternative Name | a-CTXI, aCTX-I, A-Cr, Type I Collagen C-Telopeptide-Related Fraction |
Standard | 6000pg/mL |
Sensitivity | 56.25 pg/mL |
Detection Range | 93.75-6000pg/mL |
Sample Type | Serum, Plasma, Tissue homogenate and Other biological samples |
Sample Volume | 50 μL |
Assay Type | Competitive |
Assay Duration | 1.5H |
Detection Wavelength | 450nm |
Storage | 2-8oC/-20 oC |
PRECISION
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Item | Intra-assay Precision | Intra-assay Precision |
Sample | 3 | 3 |
Replicate | 9 | 18 |
CV(%) | 5 | 8 |
Rate of recovery
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type | Range(%) | Average Recovery(%) |
Serum (n=8) | 90-105 | 97 |
EDTA plasma (n=8) | 88-102 | 95 |
Cell culture media (n=8) | 82-97 | 89 |
At sem a enim eu vulputate nullam convallis Iaculis vitae odio faucibus adipiscing urna.