Principle

This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with ACTH protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to ACTH. Next,Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of stop solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of ACTH in the samples is then determined by comparing the OD of the samples to the standard curve.

Specification

Product Name Quanticon Mouse ACTH(Adrenocorticotropic Hormone) ELISA Kit
SKU ABQ-EC96-ACTHms
Applications ELISA
Reactivity Mouse
Alternative Name Corticotropin, Adrenocorticotropin
Standard 1000pg/mL
Sensitivity 9.38 pg/mL
Detection Range 15.63-1000pg/mL
Sample Type Serum, Plasma, Tissue homogenate and Other biological samples
Sample Volume 50 μL
Assay Type Competitive
Assay Duration 1.5H
Detection Wavelength 450nm
Storage 2-8oC/-20 oC

PRECISION

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Item Intra-assay Precision      Intra-assay Precision
Sample 3 3
Replicate 9 18
CV(%) 5 8

Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

Sample Type Range(%) Average Recovery(%)
Serum (n=8) 95-107 101
EDTA plasma (n=8) 80-93 86
Cell culture media (n=8) 80-95 87