No products in the cart.
SKU: | ABQ-ES96-CXCL11ms |
Size: | Rate |
96T: | $450 |
48T: | $230 |
20 people are viewing this product right now
Principle
The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to I-TAC/CXCL11. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to I-TAC/CXCL11. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain I-TAC/CXCL11, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of I-TAC/CXCL11 in the samples is then determined by comparing the OD of the samples to the standard curve.
Specification
Product Name | “Quanticon” Mouse I-TAC/CXCL11(Interferon Inducible T-Cell Alpha Chemoattractant) ELISA Kit |
SKU | ABQ-ES96-CXCL11ms |
Applications | ELISA |
Reactivity | Mouse |
Alternative Name | CXCL11, H174, IP-9, IP9, SCYB11, SCYB9B, b-R1 |
Standard | 4000pg/mL |
Sensitivity | 37.5 pg/mL |
Detection Range | 62.5-4000pg/mL |
Sample Type | Serum, Plasma, Tissue homogenate and Other biological samples |
Sample Volume | 100 μL |
Assay Type | Sandwich |
Assay Duration | 3.5H |
Detection Wavelength | 450nm |
Storage | 2-8oC/-20 oC |
PRECISION
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Item | Intra-assay Precision | Intra-assay Precision |
Sample | 3 | 3 |
Replicate | 9 | 18 |
CV(%) | 5 | 8 |
Rate of recovery
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type | Range(%) | Average Recovery(%) |
Serum (n=8) | 92-105 | 98 |
EDTA plasma (n=8) | 90-105 | 97 |
Cell culture media (n=8) | 85-97 | 91 |
At sem a enim eu vulputate nullam convallis Iaculis vitae odio faucibus adipiscing urna.